구분
|
내용
|
Measuring gas
|
O2
|
Toxic gases(CO, Cl2, NH3 & etc) |
Measuring type
|
Auto Sampling type
|
Measuring method
|
Galvanic cell or Electrochemical cell
|
Electrochemical cell |
Measuring range
|
Refer to Toxic Gas Measuring Range
|
Response time
|
< 15sec / 90% scale
|
< 45sec / 90% scale |
Accuracy
|
±3% / Full scale
|
Parameter control
|
Front switch (calibration, maintenance, alarm setting)
|
Operation mode display
|
4-LED (Power, 2-stage alarm, 1-trouble)
|
Measuring value display
|
LCD Display(4-Digit)
|
Alarm signal output
|
2A@30VDC, 0.5A@125VAC (resistive load)
|
3-SPST relay contact(2stage alarm, 1 trouble) |
Output signal
|
Analog : 4-20mA Analog Output
|
Maintenance signal : 4mA
|
Calibration signal : 3mA
|
Fault signal : 0mA
|
Digital : RS-485 Modbus(Option) |
PoE(Power over Ethernet)(Option) |
Sampling method
|
Diaphragm Pump
|
Sampling line
|
OD 1/4" (Standard)
|
Flow rate
|
0 ~ 1000 ml/min max
|
Conduit connection
|
Standard : PF 3/4"
|
Cable gland(PG-16) |
Mounting type
|
Wall mount
|
Operating temperature
|
-10°C ~ +60°C
|
Operating humidity
|
5 to 99% RH (Non-condensing)
|
Material
|
Front(Poly carbonate), Sides(Steel)
|
Dimensions
|
Standard type : 70(W) × 144(H) × 160(D) mm
|
Pyrolyzer type(PY-1000) : 70(W) × 242.5(H) × 160(D) mm |
Weight
|
Standard type : 2.5kg
|
GTD-5000 & PY-1000 : 4.05kg |
NOTE1
|
Current consumed by Polyzer(PY1000, PY2000) Unit
|
NOTE2
|
DC 24V should be supplied if you want to use Pyrolyzer in case of using PoE power
|